Additel 672 Digital Pressure Calibrators
Price Each
Quote Only
(exc. GST)
- Description
- Overview
- Features
- Technical Specifications
- Applications
- Order Info
- Additional information
Description
? Pressure ranges to 40,000 psi (2,800 bar)
? HART Communication capability
? Measure mA or V, and with 24V loop power
? Easy-to-use, inexpensive pressure calibrator with uncertainty better than 0.02%FS
With advanced microprocessor technology and state-of-the-art silicon pressure sensors, the 672 series precision pressure calibrators provide a pressure calibration solution for gauges, transmitters, and switches over a wide pressure range. The 672 is a very portable device and can measure pressure precisely with a built-in pressure sensor, as well as read the current or mV produced by a transducer. It can even supply an excitation voltage to power sensors or transmitters during calibration. In order to reach 0.02% F. S. accuracy up to 10,000 psi (700 bar) and 0.1% F. S. accuracy up to 40,000 psi (2,800 bar), every silicon pressure sensor in our gauges has been specially aged, tested and screened before assembly. The 672 series precision pressure calibrators are unmatched in performance and reliability.
• Pressure measurement uncertainty:
ADT672-02: 0.02% FS
ADT672-05: 0.05% FS
• Pressure ranges to 40,000 psi (2,800 bar)
• Measure mA with 0.02%RD + 0.9 μA accuracy
Measure V with 0.02%RD + 0.9 mV accuracy
• Power transmitters during test using 24V loop supply
• Pressure switch test
• HART Communication capability
• Advanced temperature compensation
• Dual readout
• Min/Max/Hold to capture changing measurements
• Data logging
• Large, easy to read display with 6-digit resolution
• Backlit display
• Rechargeable battery or AC adapter
• NIST traceable calibration with data (Included)
• Accuracy
ADT672-02: 0.02% of full scale
ADT672-05: 0.05% of full scale
>20,000 psi (1,400 bar): 0.1% FS
• Gauge Types
Gauge pressure
Compound Pressure
Absolute pressure
Differential pressure
• Display
Description: Dual-line 6 full digit FSTN LCD with LED backlight
Display rate: 3.5 readings per second (Default setting)
Numeral display height: 16.5mm (0.65″)
• Pressure Units
Pa, kPa, MPa, psi, bar, mbar, inH2O@4°C, mmH2O@4°C, inHg@0°C, mmHg@0°C, kgf/cm2
• Environmental
Compensated Temperature: 14°F to 122°F (-10°C to 50°C)
Operating Temperature: 14°F to 122°F (-10°C to 50°C)
Storage Temperature: -4°F to 158°F (-20°C to 70°C)
Humidity: <95%
• Pressure Connections
1/4NPT male, 1/4BSP male, M20X1.5 male (≤15,000 psi)
1/4HP female or 1/4HP male (>15,000 psi)
*1/4HP female: Autoclave F-250-C, 9/16″ – 18 UNF-2B
*1/4HP male: Autoclave M-250-C, 9/16″ – 18 UNF-2A
0.236 inch (Ø6mm) test hose (for differential pressure)
Other connections available per request
• Electrical Connection
0.156 inch (Ø4mm) sockets
• Over Pressure Warning:
120%
• Data Logger
Storage capacity: 30 files, 40 records per file
Mode: manual and automatic
Hourly-record: record the data every hour
Interval-record: set by user
• Power
Battery: Rechargeable Li-ion polymer battery
Li-Battery working time: 40 hours
Recharge time: 4 hours
External power: 110V/220V power adapter (DC10V)
• Enclosure
Case material: Aluminum alloy
Wetted parts: 316L SS
Dimension: Ø120mm X 46mm depth X 184mm height
Weight: 0.7kg
European Compliance: CE marked
• Communication
RS232 (DB9/F, environmentally sealed)
Baud rate: 1200,2400,4800,9600
Data length: 8 bits
Stop bit: 2 bits
Address: from 1 to 112
• Warranty:
1 year
• Pressure Ranges
Gauge Pressure [1] |
|||||
P/N |
Pressure Range |
Media[2] |
Accuracy(%FS) |
Burst |
|
(psi) |
(bar) |
||||
V15 |
-15 |
-1.0 |
G |
0.02 (0.05) |
3× |
GP2 |
2 |
0.16 |
G |
0.05 |
3× |
GP5 |
5 |
0.35 |
G |
0.05 |
3× |
GP10 |
10 |
0.7 |
G |
0.02 (0.05) |
3× |
GP15 |
15 |
1.0 |
G, L[3] |
0.02 (0.05) |
3× |
GP30 |
30 |
2.0 |
G, L[3] |
0.02 (0.05) |
3× |
GP50 |
50 |
3.5 |
G, L |
0.02 (0.05) |
3× |
GP100 |
100 |
7.0 |
G, L |
0.02 (0.05) |
3× |
GP150 |
150 |
10 |
G, L |
0.02 (0.05) |
3× |
GP300 |
300 |
20 |
G, L |
0.02 (0.05) |
3× |
GP500 |
500 |
35 |
G, L |
0.02 (0.05) |
3× |
GP600 |
600 |
40 |
G, L |
0.02 (0.05) |
3× |
GP1K |
1,000 |
70 |
G, L |
0.02 (0.05) |
2× |
GP2K |
2,000 |
140 |
G, L |
0.02 (0.05) |
3× |
GP3K |
3,000 |
200 |
G, L |
0.02 (0.05) |
3× |
GP5K |
5,000 |
350 |
G, L |
0.02 (0.05) |
3× |
GP10K |
10,000 |
700 |
G, L |
0.02 (0.05) |
2× |
GP15K |
15,000 |
1,000 |
G, L |
0.05(0.1) |
2× |
GP20K |
20,000 |
1,400 |
G, L |
0.05(0.1) |
1.5× |
GP25K |
25,000 |
1,600 |
G, L |
0.1 |
1.5× |
GP30K |
30,000 |
2,000 |
G, L |
0.1 |
1.5× |
GP36K |
36,000 |
2,500 |
G, L |
0.1 |
1.5× |
GP40K |
40,000 |
2,800 |
G, L |
0.1 |
1.1× |
Note: [1]. Sealed gauge pressure for above 1000 psi
[2]. G=Gas, L=Liquid (please specify media type when place order)
[3]. 0.02% FS for gas media only
Compound Pressure |
|||||
P/N |
Pressure Range |
Media |
Accuracy(%FS) |
Burst |
|
(psi) |
(bar) |
||||
CP2 |
±2 |
±0.16 |
G |
0.05 |
3× |
CP5 |
±5 |
±0.35 |
G |
0.02 (0.05) |
3× |
CP10 |
±10 |
±0.7 |
G |
0.02 (0.05) |
3× |
CP15 |
±15 |
±1 |
G |
0.02 (0.05) |
3× |
CP30 |
-15 to 30 |
-1 to 2 |
G |
0.02 (0.05) |
3× |
CP100 |
-15 to 100 |
-1 to 7 |
G, L |
0.02 (0.05) |
3× |
CP300 |
-15 to 300 |
-1 to 20 |
G, L |
0.02 (0.05) |
3× |
Absolute Pressure |
|||||
P/N |
Pressure Range |
Media |
Accuracy(%FS) |
Burst |
|
(psi) |
(bar) |
||||
AP5 |
5 |
0.35 |
G |
0.1 |
3× |
AP10 |
10 |
0.7 |
G |
0.1 |
3× |
AP15 |
15 |
1.0 |
G |
0.1 |
3× |
AP30 |
30 |
2.0 |
G |
0.1 |
3× |
AP50 |
50 |
3.5 |
G |
0.1 |
3× |
AP100 |
100 |
7.0 |
G , L |
0.05 (0.1) |
3× |
AP300 |
300 |
20 |
G , L |
0.05 (0.1) |
3× |
AP500 |
500 |
35 |
G , L |
0.05 (0.1) |
3× |
AP1K |
1,000 |
70 |
G , L |
0.05 (0.1) |
2× |
AP3K |
3,000 |
200 |
G , L |
0.05 (0.1) |
3× |
AP5K |
5,000 |
350 |
G , L |
0.05 (0.1) |
3× |
Differential Pressure |
|||||
P/N |
Pressure Range |
Media |
Accuracy(%FS)[1] |
Burst |
|
(inH2O) |
(mbar) |
||||
DP1 |
±1 |
±2.5 |
G |
0.05[2] |
100× |
DP2 |
±2 |
±5.0 |
G |
0.05[2] |
100× |
DP5 |
±5 |
±10 |
G |
0.05[2] |
50× |
DP10 |
±10 |
±25 |
G |
0.05[2] |
20× |
DP20 |
±20 |
±50 |
G |
0.05 |
20× |
DP30 |
±30 |
±75 |
G |
0.05 |
15× |
DP50 |
±50 |
±160 |
G |
0.05 |
3× |
DP150 |
±150 |
±350 |
G |
0.02 |
3× |
DP300 |
±300 |
±700 |
G |
0.02 |
3× |
[1] FS specification applies to the span of the range. Accuracy includes one year stability.
[2] 0.05%FS accuracy (incl 6 months stability). One year accuracy is 0.05%FS
calibration accuracy combined with 0.05%FS one year stability.
Electrical Measurement and Source Accuracy |
||
Range |
Accuracy |
|
Voltage DC |
±30.0000V |
± (0.01%RD+1.5mV) |
Current DC |
±30.0000mA |
± (0.01%RD+1.5μA) |
DC 24V |
24V±0.5V max:50mA Protect at: 120mA |
|
Switch[1] |
Status OPEN/CLOSED |
Note: [1] 1V~12V if switch has detective voltage
• Gauge calibration
• Precision pressure measurement
• Pressure transmitters calibration
• Pressure switch testing
• Safety relief valve resting
• Pressure regulator testing
• Hydraulic leak testing
Additional information
Brands | Additel |
---|